tina96e1-b (2016-12) VGC501_VGC502_VGC503.om 61
DATA LOGGER >I The data logger group is used for
• recording measurement data on a USB memory
stick (interface type A on the front of the Center
unit)
• deleting recorded measurement data from the USB
memory stick
This group is only available when a USB memory stick formatted for the
FAT file system (FAT32) is plugged in. Use a max. 32 GB memory stick.
Not all USB memory sticks are automatically recognized by the
VGC50x, as they (in particular cheaper brands) do not always conform
to USB standard requirements. Try a different memory stick before
contacting your nearest INFICON service center.
DATE I Current date
TIME I Current time
INTERVAL I Display interval
DEC-SEPARATOR I Decimal separator
FILENAME I File name
START / STOP I Start / stop record
CLEAR I Deletion of files with displayed measurement data
Value
DATE I Current date in the format YYYY-MM-DD
e.g. DATE 2015-04-15 I e.g. 2015-04-15
Value
TIME I Current time in the format hh:mm [24 h]
e.g. TIME 15:45 I e.g. 15:45
Data logging interval.
Value
INTERVAL I
INTERVAL 1s I Recording interval 1/s
INTERVAL 10s I Recording interval 1/10 s
INTERVAL 30s I Recording interval 1/30 s
INTERVAL 1min I Recording interval 1/60 s
INTERVAL 1% I Recording interval: in the event of
measurement value changes ≥1%
INTERVAL 5% I Recording interval: in the event of
measurement value changes ≥5%
4.5.6 Data Logger Mode
Parameters in this group
Date
Time
Interval