tina96e1-b (2016-12) VGC501_VGC502_VGC503.om 97
Transmit: WDT [,a] <CR>[<LF>]
Description
a Watchdog control, a =
0 –> Manual error acknowledgement
1 –> Automatic error acknowledgement
1)
(default)
1)
If the watchdog has responded, the error is automatically acknowl-
edged and cancelled after 2 s.
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a <CR><LF>
Description
a Watchdog control
Transmit: AYT <CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a,b,c,d,e <CR><LF>
Description
a Type of the unit, e.g. VGC503
b Model No. of the unit, e.g. 398-483
c Serial No. of the unit, e.g. 100
d Firmware version of the unit, e.g. 1.00
e Hardware version of the unit, e.g. 1.0
Transmit: ETH [,a,bbb.bbb.bbb.bbb,ccc.ccc.ccc.ccc,ddd.ddd.ddd.ddd]
<CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a,bbb.bbb.bbb.bbb,ccc.ccc.ccc.ccc,ddd.ddd.ddd.ddd <CR><LF>
Description
a DHCP (dynamic host configuration protocol),
a =
0 –> Statically
1 –> Dynamically
bbb.bbb.bbb.bbb IP address
ccc.ccc.ccc.ccc Subnet address
ddd.ddd.ddd.ddd Gateway address