74 tina96e1-b (2016-12) VGC501_VGC502_VGC503.om
Transmit: RES [,a] <CR>[<LF>]
Description
a a =
1 –> Cancels currently active error and re-
turns to measurement mode
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: b[,b][,b][…] <CR><LF>
Description
b List of all present error messages, b =
0 –> No error
1 –> Watchdog has responded
2 –> Task fail error
3 –> FLASH error
4 –> RAM error
5 –> EEPROM error
6 –> DISPLAY error
7 –> A/D converter error
8 –> UART error
9 –> Gauge 1 general error
10 –> Gauge 1 ID error
11 –> Gauge 2 general error
12 –> Gauge 2 ID error
13 –> Gauge 3 general error
14 –> Gauge 3 ID error
RES - Reset