Service Item Estimated Service Requirement Frequency
Battery 500 full charge/discharge cycles (to approximately 80% capacity)
Compressor Function of flow setting and time
17,000 run hours @ flow setting 5
20,000 run hours @ flow setting 2
Oxygen Sensor No Service Requirement
Molecular Sieve 36 to 60 months depending on usage
Valves 20,000 run hours
Intake Filter 20,000 run hours
Cooling Fan 20,000 run hours
4. INOGEN ONE G2 SYSTEM SPECIFICATIONS
INOGEN ONE G2 CONCENTRATOR
Dimensions:
Length Width Height
Cm In Cm In Cm In
No battery
27.2 10.7 9.9 3.9 22.2 8.75
12 cell
battery
27.2 10.7 9.9 3.9 24.1 9.50
24 cell
battery
27.2 10.7 9.9 3.9 26.0 10.25
In Bag
27.2 10.7 11.4 4.50 25.4 10.00
On Cart
34.3 13.5 20.3 8.00 40.6 16.00
Weight: Approximately 7.25 pounds (includes 12 cell battery)
Noise: <38 dB on flow setting 2
Warm-Up Time: Approximately 2 minutes
Oxygen Concentration: 90 + 6/-3%
Flow Control Settings: 1,2,3,4,5
Power:
AC/DC Power Supply
AC Input: 100 to 240 VAC, 50 to 60 Hz, 1.0A to 0.5A
DC Input: 13.5 to 15 VDC, 10A
Output 19 VDC, 90 W
AMSR
oxygenconcentratorstore.com
1-877-774-9271

Other manuals for Inogen One G2