INOGEN ONE G2 OXYGEN CONCENTRATOR
TECHNICAL MANUAL
Page 18 of 20
Rechargeable Battery: Voltage: 12.0 to 16.8 VDC,
Battery Duration:
Approximately 2 to 3.5 hours for 12 cell battery
Approximately 4 to 7 hours for 24 cell battery
Battery Charging Time:
Approximately 2.25 to 4.25 hours for 12 cell battery
Approximately 4.25 to 8.00 hours for 24 cell battery
Environmental Ranges
Intended for Use:
Temperature: 40
104
o
F (4
40
o
C)
Humidity: 0% to 95%, non-condensing
Classifications
Mode of Operation: Continuous Duty
Type of Protection Against Electrical Shock: Class II
Degree of Protection Against Electrical
Shock:
Type BF
Not intended for cardiac application
Degree of Protection Against Ingress of
Water:
Concentrator: IP 20
Carry bag: IP 22
Degree of Safety for Application in Presence
of Anesthetic Gases:
Not suitable for such applications
5. INOGEN ONE G2 ERROR CODE TABLE
If more than one alarm is triggered simultaneously, the error code number will be the sum of the
error codes given below.
Error Code
Number
Explanation Possible Causes & Troubleshooting
Instructions
001 Low voltage on power supply Replace the power supply or remove
the battery to see if the concentrator
was running on battery power
002 Software error Unplug the external power and remove
the battery to clear the error and then
restart the concentrator
004 B/C Pressure Error The concentrator was shut down due
to a pressure error. Ensure that the
AMSR
oxygenconcentratorstore.com
1-877-774-9271