CONFIDENTIAL and
PROPRIETARY
96-02735-00-01 revC
INOGEN ONE G2 OXYGEN CONCENTRATOR
TECHNICAL MANUAL
Page 9 of 20
volume, and other physiologic factors. Inogen suggests that each patient be titrated
(a) while sedentary, (b) while active or ambulatory.
2.4.4. Use with the Inogen Satellite Conserver
The Inogen One G2 is not compatible with the Inogen One Satellite Conserver
3. SERVICING THE INOGEN ONE G2
3.1. Maintenance by the Provider
3.1.1. The Inogen One G2 is only intended to be repaired by authorized repair centers.
If a repair is required, please contact Inogen to locate your nearest
authorized repair center. If you would like to become an authorized repair
center, please contact Inogen for access to certification training, service
instructions, component part lists and the necessary repair equipment.
3.1.2. Checking Life Clock, Serial #, and SW version
To check usage, press and hold the Alarm Bell Button for 5 seconds. The concentrator’s
display will show the hour meter, the unit’s serial number, and the software version
installed on the device.
NOTE
Do not disassemble the Inogen One G2 or any of the accessories or attempt any
maintenance other than tasks described in this Technical Manual unless you have
completed a training course through Inogen. Disassembly of the Inogen One G2
or any of the accessories without proper training certification will void the product
warranties. Contact Inogen for information about receiving proper training and
certification for service of the Inogen One G2 and accessories.
3.1.3. Suggested Materials for Regular Maintenance
To perform regular field maintenance (by a technician) on the Inogen One G2 System,
you may need:
o Replacement cannulas (RP-128)
o Replacement intake screens (RP-200)
o Replacement Inogen One G2 Batteries (BA-200 or BA-224)
o Hose barb removal tool (Spanner Wrench, RP-102)