INOGEN ONE G2 OXYGEN CONCENTRATOR
TECHNICAL MANUAL
Page 20 of 20
and the cannula barb.
If the problem persists, return the
concentrator for service.
Battery Error The system is not properly
communicating with the battery
Remove and reinsert the battery to
clear the error. If the error persists,
replace the battery
Service Needed The compressor has reached
maximum speed
The compressor is at its end of life or
there is a leak in the concentrator
External Power
Low
The input voltage is less than 17V Remove power to the concentrator and
remove and reinsert the battery.
Try another external power source
such as DC power
6. CONTACTS FOR MORE INFORMATION
Inogen Website: Patient Issues: http://www.inogenone.com
Corporate: http://www.inogen.net
Inogen : 326 Bollay Dr
Goleta, CA 93117
1-866-765-2800
clientservices@inogen.net
AMSR
oxygenconcentratorstore.com
1-877-774-9271

Other manuals for Inogen One G2