26
Low Priority Alerts
The following low priority alert messages are accompanied by a double beep and a
solid red light.
Condition/Action/Explanation
External power supply is too low to run the unit, which
is running on battery power despite being plugged in.
Check external power supply connections. If condition
persists, contact your equipment provider.
External power supply is too high. Check external power
supply connections. If condition persists, contact your
equipment provider.
Battery power is low, with less than 10 minutes remaining.
Attach external power supply or power down and insert
a fully charged battery.
Battery error has occurred. Switch to a new battery or
remove battery and operate concentrator using external
power supply. If battery error recurs with same battery,
stop using the battery and contact your equipment
provider.
Message Display & Text
External
Power Low
External
Power High
Battery Low
Attach Plug
Battery Error
See Manual
Notifications (continued)
If multiple alarm conditions are detected, the highest priority alarm will be displayed.
The following notification messages are accompanied by a single, short beep.
Condition/Action/Explanation
On/Off button has been pressed for two seconds.
Concentrator is performing system shut down.
Audible Alarm button has been pressed for five seconds.
Message Display & Text
Inogen One®
Shutting Down
HH:MM
Vx.x:Serial Number

Other manuals for Inogen One G2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Inogen One G2 and is the answer not in the manual?

Inogen One G2 Specifications

General IconGeneral
TypePortable Oxygen Concentrator
Operating AltitudeUp to 10, 000 ft (3, 048 m)
Battery DurationUp to 4 hours (single battery)
Flow Settings1 to 5
Noise Level38 dBA
Oxygen FlowPulse Dose
Battery Duration (Double Battery)Up to 8 hours
AC Power100-240V, 50-60Hz
DC Power12V
Oxygen Concentration90%
Oxygen OutputPulse dose
Oxygen Purity90% - 96%
PowerAC/DC power supply