Inspire Sleep Remote English 27
Note: Pressing the Therapy Off button ( ) cancels
communication in progress. The sleep remote then attempts
to turn your generator off.
Pausing Therapy
If you wake with your therapy on, and you intend to go back
to sleep, you can pause therapy instead of turning therapy
off. The length of time that the therapy pauses is set by your
doctor.
Therapy Pause and Start Delay are different. Start Delay
begins automatically when you turn therapy on. When
therapy is on, Therapy Pause begins when you press the
Therapy On button ( ).
To pause therapy:
1. Press the Therapy On button ( ).
2. Position the sleep remote directly over your generator
and against your body with the buttons facing away from
your body (Figure 15).
3. Wait until the sleep remote produces a tone or 10
seconds pass.
4. Check the status ring.
– If the status ring is slowly dimming and brightening
green, therapy is paused. The status ring stays lit for
several seconds after communication with the
generator.
– If the sleep remote lights are off, gently shake the
remote and the lights will turn on.