Chapter: Requirements and Specifications
2-6 M10-16250-EN
Site Requirements for System Accessories
Some of Instron’s accessories require additional site requirements. If you are not sure if the
accessories you purchased require any of the following site preparations, contact Instron’s
Professional Services department or your local Instron office for advice.
Dry Compressed Air
Dry, compressed air is required for pneumatic grips. Typically, 90 psi air is required with a
water trap/filter.
Water Supply
A water supply may be required for some specimen grips and some temperature control
systems.
Liquid Nitrogen and Carbon Dioxide
A Liquid Nitrogen (LN2) or Carbon Dioxide (CO2) supply may be required for
environmental chambers with cooling options.
For carbon dioxide, tank pressure must be approximately 725 psi. It must be in a liquid state
when it reaches the valve of the environmental chamber.
For liquid nitrogen, tank pressure must be approximately 22 psi.