CUSA EXcel
Console Testing to IT0032 Issue 2
6. 23kHz Hand pieces output measurements
The process for the 23kHz straight and angled hand pieces are
identical to the 36kHz. However, the values are different as shown
23kHz Straight hand piece
Q Power Max Power Res power
70 Watts Max - 30 Watts Min
Frequency Between 22.8kHz and23.2kHz
23kHz Angled hand piece
Q Power Max Power Res power
45 Watts Max - 20 Watts Min
Frequency Between 22.8kHz and23.2kHz
Console Full Function Testing