Norwegian Speech Calibration
Norway TDH39
Coupler IEC 60318-3 1998 (6ccm)
Speech Audiometer
Speech Speech Noise
Norway Norway
RETSPL MaxHL RETSPL MaxHL
Linear Type A 40.0 90 40.0 80
Equ. FF Type A-E 0.5 120 0.5 115
Not Linear 7.0 120 7.0 115
Speech WN 22.5 95
Norway HDA 280
Coupler IEC 60318-3 1998 (6ccm)
Speech Audiometer
Speech Speech Noise
Norway Norway
RETSPL MaxHL RETSPL MaxHL
Linear Type A 40.0 80 40.0 75
Equ. FF Type A-E 6.5 110 6.5 105
Not Linear 7.5 120 7.5 120
Speech WN 22.5 95
Norway HDA 200
Coupler IEC60318-1 with type 1 adaptor
Speech Audiometer
Speech Speech Noise
Norway Norway
Frequency RETSPL MaxHL RETSPL MaxHL
Linear Type A 40.0 70 40.0 65
Equ. FF Type A-E 3.5 100 3.5 95
Not Linear 5.5 110 5.5 105
Speech WN 22.5 90
Norway DD45
Coupler IEC 60318-3 1998 (6ccm)
Speech Audiometer
Speech Speech Noise
Norway Norway
RETSPL MaxHL RETSPL MaxHL
Linear Type A 40.0 90 40.0 80
Equ. FF Type A-E 3.5 115 3.5 115
Not Linear 6.0 120 6.0 115
Speech WN 22.5 95
Norway EAR 3A
Coupler IEC 60318-5 2006
Speech Audiometer
Speech Speech Noise
Norway Norway
RETSPL MaxHL RETSPL MaxHL
Speech 40.0 80 40.0 70
Speech WN 22.5 85
Norway EAR 5A
Coupler IEC 60318-5 2006
Speech Audiometer
Speech Speech Noise
Norway Norway
RETSPL MaxHL RETSPL MaxHL
Speech 40.0 80 40.0 70
Speech WN 22.5 85
Norway CIR 22/CIR33
Coupler IEC 60318-5 2006
Speech Audiometer
Speech Speech Noise
Norway Norway
RETSPL MaxHL RETSPL MaxHL
Speech 40.0 70 40.0 70
Speech WN 22.5 85
Norway B71
Coupler 60318-6 2007
Speech Audiometer
Speech Speech Noise
Norway Norway
RETSPL MaxHL RETSPL MaxHL
Speech 75.0 40 75.0 30
Speech WN 57.5 55
Norway Free Field
ISO 389-7 2005 Free Field max SPL
Free Field max HL is found by subtracting the
selected RETSPL value
Binaural Monaural Free Field Power Free Field Line
45° 90° 45° 90° - 45° - 90° - 45° - 90°
RETSPL RETSPL RETSPL RETSPL RETSPL RETSPL Max SPL Max SPL
Speech 0.0 -4.0 -5.5 2.0 -2.0 -3.5 90 100
Speech Noise 0.0 -4.0 -5.5 2.0 -2.0 -3.5 85 100
Speech WN 2.5 -1.5 -3.0 4.5 0.5 -0.5 87.5 97.5

Other manuals for Interacoustics AC40