4. Maintenance
4.1 General maintenance procedures
The perf ormance and saf ety of the instrument will be kept if the f ollowing recommendations for care and
maintenance are observed:
The instrument must go through at least one annual overhaul, to ensure that the acoustical,
electrical, and mechanical properties are correct. This should be made by an authorised workshop to
guaranty proper service and repair as Interacoustics provides the necessary circuit diagrams etc. to
these repair shops.
To ensure that the reliability of the instrument is kept, it is recommended that the operator at short
intervals, for instance once a day, perform a test on a person with known data. This person could be
the operator him/herself .
After each examination of a patient, it should be ensured that there is no contamination on the parts
in connection with the patient. General precautions must be observed to avoid that disease f rom one
patient is conducted to others. If ear cushions or ear tips are contaminated, it is strongly
recommended to remove them from the transducer before they are cleaned. By frequent cleaning
water should be used, but by severe contamination it may be necessary to use a disinfectant. The
use of organic solvents and aromatic oils must be avoided.
4.2 How to clean Interacoustics products
Before cleaning always switch of f and disconnect f rom the power supply
Use a soft cloth lightly dampened with cleaning solution to clean all exposed surf aces
Do not allow liquid to meet the metal parts inside the earphones/headphones
Do not autoclave, sterilize, or immerse the instrument or accessory in any f luid
Do not use hard or pointed objects to clean any part of the instrument or accessory
Do not let parts that have been in contact with fluids dry before cleaning
Rubber ear-tips or foam ear-tips are single use components
Recommended cleaning and disinfection solutions:
Warm water with mild, nonabrasive cleaning solution (soap)
Procedure
Clean the instrument by wiping outer case with a lint free cloth lightly dampened in cleaning solution
Clean cushions and patient hand switch and other parts with a lint f ree cloth lightly dampened in
cleaning solution
Make sure not to get moisture in the speaker portion of the earphones and similar parts