INTERFACE MODEL 9840 VER 6.0.2
Disabling the window sets this value to 0. The window, once enabled and set, will take
effect immediately and will remain in effect until the window is disabled.
Note that changing a cell type on a channel will automatically disable the filter window
for that channel, reset the units to Lb or LbI, and set the window value to 0. The
window must then be reset and re-enabled by the user.
Example:
@123DW1A1
Acknowledge:
@123 Filter Window A is On
Example:
@123DW2A000.0002#
Acknowledge:
@123 Filter Window A Unit = Lb
@123 Filter Window A = 0.0002 Lb
MODEL 9840 PG 50 PUB. 2856-16