Circulating Baths with
Standard Digital Temperature Controller
Operator’s Manual
110-512 PSC/EN 28 May 2014
Models:
SD07R-20
SD7LR-20
SD15R-30
SD20R-30
SD28R-30
SD07H170
SD15H170
SD20H170
SD28H170
SD29VB3S
SD29VB5R

Table of Contents

Questions and Answers

  • A
    Angela RussellAug 13, 2025
    Why is my Intertek SD07R-20 Controller not cooling enough?
    • L
      Lindsey BenjaminAug 13, 2025
      If the Intertek Controller isn't cooling sufficiently, several factors could be at play. Begin by ensuring the air filter and condenser are free from dust, and that the air ventilation screens are not blocked. Verify that the temperature set point isn't too high and that the heat load doesn't exceed the bath's capacity. The ambient air temperature should also be checked to ensure it's not too high (greater than 35°C / 95°F). Finally, check for any issues with the line voltage and correct them if necessary.
  • W
    Walter RobertsAug 16, 2025
    What to do if Intertek SD07R-20 Controller does not cool?
    • C
      clinejeffreyAug 16, 2025
      If your Intertek Controller isn't cooling, start by checking for dust buildup on the air filter or condenser and clean them if needed. Also, make sure the air ventilation screens aren't blocked. Ensure the temperature set point isn't too high. Excessive heat load or a high ambient air temperature (above 35°C / 95°F) can also prevent cooling. Lastly, check and correct any issues with the line voltage.
  • O
    ortizrichardAug 20, 2025
    Why is my Intertek SD07R-20 Controller not heating up?
    • J
      Jason ReyesAug 20, 2025
      If your Intertek Controller is not heating, it could be due to several reasons. First, ensure there is sufficient fluid in the reservoir. If the fluid level is adequate, check that the temperature set point and the Safety Set Temperature are appropriately set; increase them if they are too low.
  • K
    Katherine WarrenAug 25, 2025
    How to fix insufficient heating on Intertek SD07R-20 Controller?
    • D
      Daniel AllenAug 25, 2025
      To address insufficient heating in your Intertek Controller, first, check for insufficient circulation. Also, verify and correct low line voltage if present. Ensure the ambient temperature is not too cool, or relocate the unit to a warmer environment. Lastly, inspect for any excessive heat loss from external tanks and hoses, as well as vapor/heat loss from the internal reservoir.
  • L
    Lisa HarrellAug 27, 2025
    How to turn on Intertek SD07R-20 when it shows STANDBY?
    • J
      Jared ShortAug 27, 2025
      If your Intertek Controller displays 'STANDBY' on the digital display and doesn't run, it is likely in Standby mode. To resolve this, simply press the Power Key on the front panel.
  • B
    brittany50Sep 1, 2025
    What to do if Intertek Controller has no fluid circulation?
    • R
      Rose LongSep 1, 2025
      If your Intertek Controller is not circulating fluid, it might be due to insufficient fluid in the reservoir. Try adding fluid to the reservoir. It could also be that the pump impeller is jammed. Inspect the pump and remove any debris as necessary.
  • E
    ecrawfordSep 4, 2025
    How to troubleshoot a blank digital display on Intertek SD07R-20 Controller?
    • A
      Allison NolanSep 4, 2025
      If the Intertek Controller is not running and the digital display is blank, the issue is likely a lack of power to the unit. Check that the electrical cord is securely connected to both the unit and a functioning electrical outlet.
  • B
    Brittany ScottSep 8, 2025
    Why is there insufficient circulation in my Intertek SD07R-20 Controller?
    • K
      Karen FosterSep 8, 2025
      If you're experiencing insufficient circulation with your Intertek Controller, it could be due to several reasons. The fluid viscosity might be too high, in which case you should replace it with a lower viscosity bath fluid. Another possibility is that the external tubing diameter is too small; replace it with larger diameter tubing. Low line voltage can also cause this issue, so check and correct it as needed.

Summary

Introduction to Circulating Baths

General Safety Information

Provides essential safety guidelines and warnings for operating the circulating bath safely.

Safety Recommendations

Details specific safety advice, including fluid compatibility and electrical precautions.

Quick-Start Guide

Installation and Startup Procedures

Electrical Power Connection

Guidance on safely connecting the circulator to a grounded power source.

Controller Setup and Operation

Setting Safety Temperature

How to set the safety thermostat to prevent over-temperature conditions.

Normal Operation and Controls

Controller Interface and Display

Understanding the controller's keys, display, and main operational screen.

Adjusting Temperature Set Point

Step-by-step guide to setting the desired operating temperature for the bath fluid.

Advanced Settings and Calibration

Display Messages and Troubleshooting

Display Messages and Alarms

Interpreting system messages, warnings, and alarm conditions.

Routine Maintenance Procedures

Guidelines for maintaining bath water clarity and draining the reservoir.

Troubleshooting Common Issues

Diagnosing and resolving operational problems like no circulation or temperature instability.