EasyManua.ls Logo

Intertek SD07R-20 - Setting Temperature Limits

Intertek SD07R-20
51 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibrating Your Circulator
This sub-menu allows you to match the Circulator’s temperature display to an external reference
thermometer. A value from -3.0° to +3.0°C may be entered; the factory default is 0.0°C.
IMPORTANT: To prevent Offset Calibration value from being changed unintentionally, the following
power down/power up sequence is required to enable the Calibration function.
1. Place the power switch/circuit breaker on the rear of the unit in the OFF position.
2. Return the power switch/circuit breaker to the ON position while pressing and holding the
key.
3.
When STANDBY appears on the display, release the
key and press
. You can now
proceed as outlined below.
The Calibration function will remain enabled until the electrical power is turned OFF by pressing the
key.
To Access: Press the
key until CALIBRATE appears on the display.
To Change: Touch the arrows until the desired calibration temperature is displayed.
To Accept: Press
,
, ,
or allow the LCD to return to the main operational
display (approximately 10 seconds).
CAUTION: The Offset Calibration value is always shown in degrees C, even if degrees F is selected as
the temperature unit in which the control and actual bath temperatures are displayed. Your Circulator
will automatically convert the °C offset calibration value to the correct °F display offset value.
Touch top arrow to
increase calibration offset
Touch bottom arrow to
decrease calibration offset
110-512 PSC/EN
25

Table of Contents