Command Format Values
Return
Message
Set Command Echo SEi[CR]
Echo: i = 1
No Echo: i = 0
![CR]
Set Set Point SSiii.i[CR] i = any integer from 0-9 ![CR]
Set On Off SOi[CR]
On: i = 1
Off: i = 0
![CR]
Set High Alarm SHiii[CR] i = any integer from 0-9 ![CR]
Set Low Alarm SLiii[CR] i = any integer from 0-9 ![CR]
Set Pump Speed SMi[CR]
Low: i = 1
High: i = 2
![CR]
Read Set Point Temperature RS[CR] iii.i[CR]
Read Units of Operation RU[CR] C[CR] or F[CR]
Read Internal Temperature RT[CR] iii.i[CR]
Read Operating Status RO[CR]
Running: i = 1
Standby: i = 0
i[CR]
Read High Alarm Setting RH[CR] iii[CR]
Read Low Alarm Setting RL[CR] iii[CR]
Read Pump Speed RM[CR]
Low: i = 1
High: i = 2
i[CR]
Read Alarm Status RF[CR]
No Faults: i = 0
Fault: i = 1
i[CR]
Read the Auto-Cool Set Point RA[CR] ii[CR]
Read the Firmware Version RB[CR] viiii[CR]