Using the Electrosurgical Unit (ESU) with Monopolar and Bipolar Instruments
da Vinci SP Instruments and Accessories
58
DRAFT/PRE-RELEASE/CONFIDENTIAL
12/5/19
CAUTION: Connect the ESU to the EndoWrist SP instrument using the appropriate
monopolar/bipolar cord. Refer to the ESU’s manual for indications and instructions
in making this connection. Monopolar cords may only be connected to monopolar
ports, and bipolar cords to bipolar ports.
Device Description
The reusable Bipolar Cautery Cord (PN 470384) is designed to be used as an accessory to the
EndoWrist SP bipolar instruments in conjunction with the VIO dV generator during surgical
and other medical procedures.
Figure 8.20 Bipolar cautery cord connectors
Inspection
Inspect cords for damage or irregularities before and after each use. Visually inspect the cords
for physical damage including:
• Cracked, broken, or otherwise distorted parts
• Broken or significantly bent connector contacts
• Punctures, cuts, nicks, abrasions, unusual bumps, or significant discoloration
• Corrosion of connector contacts
Wide connector
Instrument connector