EasyManua.ls Logo

Invacare Arrow - Page 126

Invacare Arrow
128 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12 TROUBLESHOOTING
Invacare
3G Storm Series® 126 Part No 1143151
Battery indicator flashes the charge level is
low - immediately after recharge.
Battery failure.
Malfunctioning battery charger.
Electrical malfunction.
Check batteries for shorted cell. Replace if necessary.
Contact Dealer/Invacare for Service.
Contact Dealer/Invacare for Service.
Battery indicator flashes the charge level is
low - too soon after being recharged.
Batteries not charged.
Weak batteries.
Have charger checked.
Replace batteries if necessary. Refer to Replacing Batteries
on page 94.
Motor “chatters” or runs irregular. Motor/gearbox malfunction. Stop use of Wheelchair. Contact Dealer/Invacare for Service.
Joystick erratic or does not respond as
desired.
Damaged motor coupling.
Electrical malfunction.
Controller programmed
improperly.
Contact Dealer/Invacare for Service.
Contact Dealer/Invacare for Service.
Contact Dealer/Invacare to have controller reprogrammed.
Wheelchair does not respond to commands. Electrical malfunction. Contact Dealer/Invacare for Service.
SYMPTOM PROBABLE CAUSE SOLUTIONS

Table of Contents

Other manuals for Invacare Arrow

Related product manuals