3) TDH39, DD45 and DD65 transducers with 2 m double signal shielded
cable
4) Insert earphone: ER-3 (Etymotic) or IP30 (RadioEar)
5) Bone vibrator B-71 transducer with twisted not shielded double signal
cable 2m
6) Patient response switch (manufactured by Inventis) with 2 m shielded
cable
7) Talk over microphone with 2m shielded cable
8) Stereo cable 3.5mm plug to 3.5mm plug, 1.8m, shielded
9) USB cable, shielded, maximum length: 2 m
The use of accessories, transducers and cables other than those
specified, except for transducers and cables sold by the manufacturer
as spare parts for internal components, may result in increased
emissions or decreased immunity of the device.
Portable RF communications equipment (including peripherals such
as antenna cables and external antennas) should be used no closer
than 30 cm (12 inches) to any part of Piccolo, including cables
specified by the manufacturer. Otherwise, degradation of the
performance of this equipment could result.
Anyone connecting additional equipment is responsible for making sure the
system complies with the IEC 60601-1-2 standard.
The instrument has no ESSENTIAL PERFORMANCE as related to IEC
60601-1
Note: All necessary instruction for maintaining compliance with regard to
electromagnetic compatibility can be found in the maintenance section in this
manual. No further steps are required.
Guidance and manufacturer’s declaration – electromagnetic emissions
Piccolo is intended for use in the electromagnetic environment specified below.
The customer or the user of Piccolo should assure that it is used in such an
environment.
Electromagnetic environment-
guidance