Piccolo uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and not likely to
cause any interference in nearby
electronic equipment.
Piccolo is suitable for use in professional
healthcare facility environment and
directly connected to the public low-
voltage power supply network.
Harmonic
emissions
IEC 61000-3-2
Voltage
fluctuations /
flickers emissions
IEC 61000-3-3
Guidance and manufacturer’s declaration – electromagnetic immunity
Piccolo is intended for use in the electromagnetic environment specified below.
The customer or the user of Piccolo should assure that it is used in such an
environment.
Electromagnetic
Environment -
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV,
± 8 kV, ± 15 kV
air
± 8 kV contact
(1)
± 2 kV, ± 4 kV,
± 8 kV, ± 15 kV
air
(1)
Floors should be
wood, concrete or
ceramic tile. If
floors are covered
with synthetic
material, the
relative humidity
should be at least
30%
Electrical fast
transient / burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input /
output lines
± 2 kV for power
supply lines
± 1 kV for input /
output lines
Mains power
quality should be
that of a
professional
healthcare facility
environment.
± 1 kV
differential mode
±2 kV common
mode
± 1 kV
differential mode
±2 kV common
mode
Mains power
quality should be
that of a
professional
healthcare facility
environment.
Voltage dips,
short
interruptions and
voltage variations
< 5% U
T
(2)
(> 95% dip in U
T
)
for 0,5 cycle.
40% U
T
< 5% U
T
(2)
(> 95% dip in U
T
)
for 0,5 cycle.
40% U
T
Mains power
quality should be
that of a
professional
healthcare facility
environment. If the