31
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular / Cordless)
telephones and land mobile, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey should be considered. If
measured field strength in the location in which Piccolo is used exceeds the applicable RF
compliance level above, Piccolo should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or
relocating Piccolo.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Guidance and manufacturer’s declaration electromagnetic immunity
Function to verify for freedom from
unacceptable risk
Acceptance pass/fail criteria
Sound generation operating correctly
No sound from trasnducers exceeding 80dB;
a lock or reboot of the device is acceptable
The Areo version contains a transmission module with FCC ID:
QOQBLE112
The Areo version contains a transmission module with IC: 5123A-
BGTBLE112
The Aereo version of the device emits radio frequencies in the 2.4 GHz
band, class 1
Contains transmitter module in compliance with EN 301 489-1 and EN
300 328 standards