EasyManua.ls Logo

IPS Controllers M820 - Sensor Verification Checks

IPS Controllers M820
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
h. Turn on the M820 controller.
i. Open the ow cell valves and wait for a few minutes for
the system to stabilize and get an accurate reading. Adjust
the Set Level if necessary.
3. Checking the ORP sensor
a. Place the sensor in a cup of tap water. The reading should
gradually drop down to 400mV.
b. Add a pinch of Dichlor, Trichlor or a drop of Liquid Chlorine
into the water. The reading should increase above 750mV.
4. Checking the pH sensor
a. Place the sensor in a cup of tap water.
b. Add a small amount of acid to the water and take a reading.
It should show a low number (probably 6.0).
c. Rinse the sensor, then place into a soluon that is higher
than 7.5pH. Verify that the reading is increasing.
M820 pH/Dual ORP Controller Owners Manual
27

Related product manuals