19
(2) Auto-V
Input Protection : 1000V
DC
or 1000V
AC
rms
AC Frequency Response : 50Hz ~ 1KHz
Input Impedance : Approx. 3kΩ.
AC Conversion Type :
Conversion type and additional specication are same as voltage function.
(3) Current
Input Protection : Protected by high energy fuses.
440mA, 1000V AC/DC 10KA Fast acting fuse for mA input. 11A, 1000V AC/DC 20KA Fast
acting fuse for A input
Max. Measuring Time : 3 minutes for “A” input, 10 minutes for “mA” input.
Rest time 20 minutes min.
AC Conversion Type : Conversion type and additional specication are same
as voltage function.
AC Frequency Response : 50Hz ~ 1KHz
IDM98-III/IDM99-III
Range Resolution IDM98-III Accuracy IDM99-III Accuracy
600.0V
DC
0.1V
± (1.0% + 3d) ± (0.80% + 3d)
1000V
DC
1V
600.0V
AC
0.1V
1000V
AC
1V
IDM98-III/IDM99-III
Range Resolution IDM98-III Accuracy IDM99-III Accuracy
60.00mA
DC
0.01mA ± (1.0% + 3d) ± (0.8% + 3d)
600.0mA
DC
0.1mA ± (1.0% + 3d) ± (0.8% + 3d)
6.000A
DC
0.001A ± (1.0% + 3d) ± (0.8% + 3d)
10.00A
DC
0.01A ± (1.0% + 3d) ± (0.8% + 3d)
60.00mA
AC
0.01mA ± (1.5% + 3d) ± (1.2% + 3d)
600.0mA
AC
0.1mA ± (1.5% + 3d) ± (1.2% + 3d)
6.000A
AC
0.001A ± (1.5% + 3d) ± (1.2% + 3d)
10.00A
AC
0.01A ± (1.5% + 3d) ± (1.2% + 3d)