(7) Diode Test
Input Protection : 1000V
DC
or 1000V
AC
rms
Max. Open Circuit Voltage : Approximately. 2.5V
Max.Test Current : Approximately 0.4mA
(8) Capacitance
Input Protection : 1000V
DC
or 1000V
AC
rms
Max. Measuring Time : 0.7 sec for 1nF~1mF
3 sec for 1mF~10mF
(9) Frequency Counter
Input Protection : 1000V
DC
or 1000V
AC
rms
Min. Frequency : 1Hz
Sensitivity : > 5.0Vp-p (for ACV 1Hz ~ 10kHz). > 10Vp-p (for ACV 10kHz ~ 100kHz).
2mAp-p (for AC mA). 0.2Ap-p (for ACA).
IDM98-III/IDM99-III
Range Resolution Accuracy
2.000V 1mV ± (1.5% + 2d)
IDM98-III/IDM99-III
Range Resolution Accuracy
1.000µF 0.001µF
± (1.2% + 2d)
10.00µF 0.01µF
100.0µF 0.1µF
1.000mF 0.001mF
10.00mF 0.01mF
IDM98-III/IDM99-III
Range Resolution Accuracy
100.00Hz 0.01Hz
± (0.1% + 2d)
1000.0Hz 0.1Hz
10.000KHz 0.001KHz
100.00KHz 0.01KHz