Dual Furnace Models M17705, N17706 & 17707 Iss.09 12/14
MODEL 17707 WATER HEAT-PIPE FURNACE
FINAL PART OF THE COMMISSIONING INSTRUCTIONS
The heat up rate of the furnace is limited to ensure long life of the heat pipe.
Impose a heat pipe set point of 150°C and set the over-temperature controller to 200°C.
Allow 1 to 2 hours for the furnace to stabilise and then reduce the over-temperature cut-out setting until the relay
trips, thus switching off the furnace.
The 17707 furnace is now ready for use.