Dual Furnace Models M17705, N17706 & 17707 Iss.09 12/14
SPECIFICATION
Dimensions of the cabinet containing the apparatus are:
Width: 600mm
Depth: 560mm
Height: 960mm (excluding thermometer storage rack)
Control: The heat-pipe and the thermometer furnace are controlled independently to a resolution of
0.1°C.
Communications: RS422
Over-temperature Provided on each furnace and operated by an independent sensor.
Protection:
Temperature range: 450 to 1100°C.
Power: 3kW, 220-240V, 50/60Hz
(110-120V to special order)
Performance: Essentially temperature-gradient-free heat-pipes provide the ideal apparatus for fixed point
calibration and give a virtually perfect freezing profile within the fixed point cell.
The addition of the second furnace with its air-flow enables thermometers to be easily and
safely calibrated.
A comprehensive manual accompanies each delivery.