inquiry syntax:[SOURce:]APPLy?
Return parameter:<NRf>,<NRf>
4.7 Input Measuring Commands
MEASure[:SCALar]:VOLTage[:DC]?
This command is used to enquire the actual output voltage.
Command syntax:MEASure[:SCALar]:VOLTage[:DC]?
Parameter:None
Return parameter:〈NRf〉
Return parameter’s unit:V
example:MEAS:VOLT?
FETCh:VOLTage?
This command is used to read the voltage to be handled in the sample cache. After sending
the command and let the instrument dialogue, the readings will be sent to the computer. This
command does not affect the instrument settings. This command does not trigger a
measurement operation, requires only the latest available reading. Before reading the new,
the command returns are old readings.
Command syntax:FETCh:VOLTage?
Return parameter:<NRf>
Return parameter’s unit:V
MEASure[:SCALar]:CURRent[:DC]?
This command is used to read the actual current.
Command syntax:MEASure[:SCALar]:CURRent[:DC]?
parameter:None
return parameter:〈NRf〉
return parameter’s unit:A
example:MEAS:CURR?