EasyManua.ls Logo

ITech IT6512 - Output Command

ITech IT6512
41 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
source mode.The function is the same as command *TRG.
Command syntax:TRIGger[:IMMediate]
parameter:None
relative commands:*TRG TRIG:SOURce
TRIGger:SOURce
This command is used to select the trigger source.Power supply could receive trigger signal
directly from front panel by pushing “Trigger” button or receive BUS trigger signal.
Command syntax:TRIG:SOURce <mode>
parameters:MANUAL |BUS
inquiry syntax:TRIGger:SOURce?
Return parameter:MANUAL |BUS
4.5 Output Command
[SOURce:]OUTPut[:STATe]
This command is used to control the output state of power supply.
Command syntax:[SOURce:]OUTPut [:STATe] <bool>
parameter:0|1|ON|OFF
inquiry syntax:[SOURce:]OUTPut[:STATe]?
Return parameter:0|1
[SOURce:]RISe[:LEVel]
This command is used to set the voltage rising time of power supply.
Command syntax:[SOURce:]RISe[:LEVel] <NRf>
unit:s
parameter:0~65.535
inquiry syntax:[SOURce:]RISe[:LEVel]?
Return parameter:0~65.535