parameter:0|1|ON|OFF, ‘6512’
example:CAL:SEC 0, ‘6512’; CAL:SEC 1
inquiry syntax:CALibration:SECure?
parameter:None
CALibration:VOLTage:LEVel
This command is used to define the voltage calibration point.P1 and P2 calibration points
should be calibrated according to priority.
Command syntax:CALibration:VOLTage:LEVel <point>
parameter:P1|P2
CALibration:VOLTage [:DATA] {<numeric value>}
This command is used to read the actual voltage of the present calibration point.
Command syntax:CALibration:VOLTage [:DATA] <NRf>
parameter:<NRf>
example:CAL:VOLT 80.0002V
CALibration:CURRent:LEVel
This command is used to define the current calibration points. P1 and P2 calibration points
should be calibrated according to priority.
Command syntax:CALibration:CURRent:LEVel <point>
parameter:P1|P2
CALibration:CURRent [:DATA] {<numeric value>}
This command is used to read the actual current of the present calibration point.
Command syntax:CALibration:CURRent [:DATA] <NRf>
parameter:<NRf>
example:CAL:CURR 3.002A
CALibration:SAVe