ITech IT6512 - USB Remote Control Interface

ITech IT6512
41 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
This command is used to savae the calibration parameters in a nonvolatile memory.
Command syntax:CALibration:SAVe
parameter:None
CALibration:STRing
This is used to set the calibration information
Command syntax:CALibration:STRing <parameter>
Parameter: The max length is a chart with 20 letters,i.e the relative information during the
calibration process,such as calibration date.
example:CAL DATE: 2011/01/01
CALibration:STRing?
This is used to view the information of the time.
Inquiry syntax:CALibration:STRing?
Return parameter:calibration information saved in the power supply
CALibration: INITialize
This command is used to initialize the calibration parameters.
Command syntax:CALibration: INITialize
parameter:None