Function and Features 
  Copyright ©ITECH Electronic Co., Ltd.  32 
The 10V reference voltage output by the power supply can 
be connected to a resistance subdivision for analog 
quantity control. 
Program 
(Current 
Setting) 
Output current of analog quantity control: 
In setting the V-PRG and 10v/10k-
quantity range should be 0-10V voltage, and the regulated 
output current should be from 0 to full range; 
In  setting the V-PRG and 5v/5k-
quantity range should be 0-5V voltage, and the regulated 
output current should be from 0 to full range; 
In setting the R-PRG and 10v/10k-
quantity range should be 0-
regulated output current should be from 0 to full range;   
In setting the R-PRG and 5v/5k-
quantity range should be 0-
regulated output current should be from 0 to full range;
Monitor 
(Voltage 
monitoring) 
The actual value from monitoring is the corresponding 
monitor voltage value. For example, when the analog 
quantity voltage is 10V, IT6512A control voltage 0~80V 
and the IT6512A output voltage 20V, this pin will output 
2.5V voltage. Similarly, when the analog quantity voltage 
is 5V, control voltage 0~80V and the IT6512A output 
voltage 20V, this pin will output 1.25V voltage. 
Monitor   
(Current 
monitoring) 
The actual value from monitoring is the corresponding 
monitor voltage value. 
For example, when the analog 
quantity voltage is 10V,  IT6512A control current 0~60A 
and the IT6512A output current 6A, this pin will output 1V 
voltage. Similarly, when the analog quantity voltage is 5V, 
control current 0~60A and the IT6512A output voltage 6A, 
this pin will output 0.5V voltage. 
11, 12 and 13 
to AGND 
Ground wires for analog quantity interfaces (including Pin 
20 VPRG, Pin 21 REF_10V, Pin 22 IPRG, Pin 23 VMON, 
Pin 24 IMON). 
connection 
Ground wires for Pin 14 EXT ON, Pin 15 SHUT OFF and 
Pin 16 POWER OK. 
Voltage setting (Voltage Program) 
This function enables change of voltage output through external analog signal by 
connecting external DC voltage (under voltage mode) or an external  resistor (under 
resistor mode) to Pin 20. To enable this function, the output control should be under the 
external analog control mode. Used for controlling the external voltage range of full-scale 
output voltage or the resistor can be selected from 0~10V/0~10KΩ or 0~5V/0~5KΩ.  To 
switch on the voltage setting, open the MENU (Menu) →  CONFIG  (Configure)  → 
Ext-Ctrl (External Analog Control). Press the Right Key to select V-P (or R-P). Press the 
Up/Down Key to select the voltage mode or resistor mode. 
  Voltage Mode 
Ext-Ctrl  Setup 
5v-M  5v/5k-P  V-P  off 
 
Under voltage mode, the user can set the voltage output value of power supply through Pin 
20.