5101015-01 Rev 2 8
2.2 Connections
Compressed Air Connection
Compressed air, nitrogen or carbon dioxide is supplied to the power
unit through a 1/4" NPT female connector. The connector contains
a screen to stop large debris from entering the unit. Do not remove
this connector. Support the connector with a wrench while installing
the user supplied compressed air fitting. Connect the compressed
air using appropriate tubing and fittings, the flow diameter of the
tubing and fittings should not be less then 3/16" (4 mm). If tubing or
fittings with inadequate flow diameter are used, performance will be
reduced. The compressed air must be clean, dry and oil free. The
maximum allowable input pressure is 100 psi (7 bar). Use of an
external pressure regulator and filter / dryer is recommended.
Power Connection
Top Gun accepts AC input of 120 VAC or 230 VAC 50/60 Hz single
phase. Top Gun must be grounded for safe and proper operation.
Top Gun is available with different line cords to meet the main
power connection plug requirements in many areas of the world.
Connect the supplied power cord to an appropriate 3-terminal
grounded AC power receptacle.
2.3 Mounting the Gun Hanger
The gun hanger can be mounted on any
convenient surface such as a wall, side of a table
top or work bench leg. The hanger can also be
mounted on the front of the power unit using the
pre-punched holes. Secure hanger with screws
and washers:
Figure 4. Top Gun Air
Gun Hanger