EasyManua.ls Logo

Ivoclar Vivadent Programat P310 - 9. Product Specifications; 9.1 Delivery Form; 9.2 Technical Data

Ivoclar Vivadent Programat P310
64 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
61
9. Product Specifications
9.1 Delivery form
– Programat P310
– Power cord
– Vacuum hose
– Programat Firing Tray Set 2
– Temperature Checking Set 2
– USB download cable
– Operating Instructions
Recommended accessories
– Programat Accessories Set
– Temperature Checking Set 2
– Vacuum Pump VP5, VP3 easy
9.2 Technical data
Power supply 110 – 120 V / 50 – 60 Hz
200 – 240 V / 50 – 60 Hz
Overvoltage category II
Contamination level 2
Acceptable voltage fluctuations ± 10 %
Max. power consumption 12 A at 110 – 120 V
8.5 A at 200 – 240 V
Acceptable data for pumps of other manufacturers:
Max. output:
Final vacuum:
250 W / max. leakage current 0.75 mA
< 50 mbar
Only tested pumps should be used
Electrical fuses
110 – 120 V:
125 V / T15A (heating circuit)
125 V / T5A (vacuum pump)
200 – 240 V:
250 V / T8A (heating circuit)
250 V / T3.15A (vacuum pump)
Dimensions of electrical fuses 110 – 120 V: Diameter 6.3 x 32 mm
200 – 240 V: Diameter 5 x 20 mm
Dimensions of the closed furnace Depth: 465 mm
Width: 320 mm / 390 mm (with cooling tray)
Height: 320 mm
Usable size of the firing chamber Diameter: 90 mm
Height: 80 mm
Max. firing temperature 1200 °C
Weight 15.6 kg
Safety notes
The furnace complies with the following standards:
– UL 61010-1:2004
– CAN/CSA-C22.2 No. 61010-1:2004
– IEC 61010-2-010:2003
– EN 61010-2-010:2003
– CAN/CSA-C22.2 No. 61010-2-010:2004
– IEC 61010-1:2010
– EN 61010-1:2010
– IEC 61010-1:2001
– EN 61010-1:2001
– UL 61010-1:2012
– CAN/CSA-C22.2 No. 61010-1:2012
Radio protection, electromagnetic compatibility: EMC tested