SPECIFICATIONS
SPECIFICATIONS
ECG
Lead Selection: LI, LII, LIII menu selectable.
Patient Cable: 6-Pin AAMI Standard connector
Isolation: Isolated from ground related circuits by >4 kV rms, 5.5 kV peak
CMRR: 90 dB with patient cable and 51 k/47 nF imbalance
Input Impedance: 20 M at 10 Hz with patient cable
Frequency Response
LCD: Filtered: 0.5 to 25 Hz
Unfiltered: 0.5 to 45 Hz
Frequency Response
X1000 output: Filtered: 0.5 to 25 Hz
Unfiltered: 0.2 to 100 Hz
Input Bias Current: Any lead configuration <200 nA dc maximum
Electrode Offset Potential: ±0.5 V DC
Lead Off sensing current: 56nA
Noise: <20 µV peak-to-peak, referred to the input with all leads connected
through 51 k/47 nF to ground
Defibrillator Protection: Protected against 360 J discharge and electrosurgery potentials
Recovery time <6s
Leakage Current: <10 µA at normal condition
Electrosurgical Interference
Protection: Standard. Recovery time: <6 seconds.
Notch Filter: 50/60 Hz (automatic).
Cardiotach
Range: 15 to 300 bpm
Accuracy: ±1%
Resolution: 1 bpm
Sensitivity: 300 µV peak
Heart Rate Averaging: Exponential averaging calculated once a second with a maximum
response time of 8 seconds.
Response Time: 8 seconds
Tall T Wave Rejection: Rejects T waves R wave
Model 3000 Series Service Addendum
25

Other manuals for Ivy Biomedical Systems 3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ivy Biomedical Systems 3000 Series and is the answer not in the manual?

Ivy Biomedical Systems 3000 Series Specifications

General IconGeneral
BrandIvy Biomedical Systems
Model3000 Series
CategoryMonitor
LanguageEnglish

Related product manuals