SPECIFICATIONS
Recorder (optional)
Writing Method: Direct Thermal
Number of Traces: 1
Modes: Direct - Manual Recording
Timed - Print button initiates a 30 second recording
Delay - Recorder is activated by an alarm
Paper Speeds: 25 and 50 mm/s
Resolution: Vertical - 200 dots/in.
Horizontal - 600 dots/in. at 25 mm/s
400 dots/in. at >25 mm/s
Frequency Response: >100 Hz at 50 mm/s
Data Rate: 400 samples/s/trace
Synchronized Output (Trigger)
Phase Delay from Electrode
Leads to Trigger Output: < 3 ms
R to R Trigger Accuracy: ±50µs typical @ 1 mV input
Pulse width: 1 to 150ms (standard 150ms)
Pulse amplitude: 0 to 5V or ± (0 to 12V)
Output Impedance: <100
Sensitivity and Threshold
Adjustment: Fully Automatic
Real Time Clock
Resolution: 1 minute
Display: 24 hours
Power Requirement: The real time clock keeps time if the monitor has power or not.
The clock is powered by a dedicated battery whose life is minimum
4 years at a temperature of 25
o
C
Environmental
Operating Temperature
Range: 5°C to 35°C
Storage Temperature
Range: -5°C to 60°C
Relative Humidity: 0-90% non-condensing
Atmospheric Pressure: 500-1060 mbar
Power Requirements
Voltage Input: 100 to 230V~
Line Frequency: 47 to 63 Hz
Fuses Type and Rating: T.5A, 250V (Metric 5x20mm)
Maximum ac Power
Consumption: 35 VA
Regulatory
Unit meets or exceeds the specifications for the AAMI Cardiac Monitor Standard EC-13, UL2601-1,
CAN/CSA C22.2 No 601.1-M90, CDN MDR (CMDCAS), IEC 60601-2-25, IEC 60601-2-27,
MDD.93/42/EEC, CE 0143, ISO 13485:1996, and FDA/CGMP.
Model 3000 Series Service Addendum
27

Other manuals for Ivy Biomedical Systems 3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ivy Biomedical Systems 3000 Series and is the answer not in the manual?

Ivy Biomedical Systems 3000 Series Specifications

General IconGeneral
BrandIvy Biomedical Systems
Model3000 Series
CategoryMonitor
LanguageEnglish

Related product manuals