4.1 Autoclaves type «B»
Autester ST DRY PV autoclaves allow sterilization of:
• Solid loads resistant to temperature: instrumental, metallic.
• Solid loads sensitive to temperature: plastics
• Solid packed load.
• Porous load: textiles
• Wrapped porous load.
• Catheters, tubes and cavities devices.
• Liquids.
4.2 Technical specications
Model 4001738 4001739
Supply voltage: 220V, 50Hz 220V, 50Hz
Power: 1500VA 1700VA
Fuse: F12AL F12AL
Working temperature: 5-40ºC 5-40ºC
Noise: < 50dBA < 50dBA
Maximum weight per tray: 1 Kg 1 Kg
Maximum heat radiation (from 20 to 26ºC): < 2000J < 2000J
Chamber material: Stainless steel Stainless steel
Maximum working pressure: 2.5bar 2.5bar
Minimum working pressure: -0.9bar -0.9bar
Maximum temperature: 145ºC 145ºC
Chamber diameter measurements: 245x360mm 245x470mm
Water volume per cycle: 0.16L 0.18L
Water tank volume: 3.5L 4L
4.3 Sterilization cycle
1-2 pre-heating
2-3 pre-vacuum
3-4 sterilizing
4-5 air-discharging
5-6 drying
6-7 stabilizing
1-7 entire duration
Sterilization cycles types
Type B: The sterilization of all wrapped or non-wrapped, solid, hollow load products type A and porous pro-
ducts as represented by the test loads in this standard.
Type S: The sterilization of products as specied by the manufacturer of the sterilizer including non-wrapped
solid products and at least one of the following: porous products, small porous items, hollow load
products type A, hollow load products B, single wrapped products, multiple-layer wrapped products.
NOTE 1 The description identies ranges of products and test loads.
NOTE 2 Non wrapped sterilized instruments are intended either for immediate use or for non sterile
storage, transport and application (e.g. to prevent cross infection).