Light Source | 150W Xenon lamp |
---|---|
Detector | Photomultiplier tube |
Model | J-715 |
Wavelength Range | 163-900 nm |
Measurement Modes | CD, Absorbance, Fluorescence |
Temperature Control | Optional Peltier temperature controller |
Software | Spectra Manager |
Verify N2 flow is 12-20l/min and cooling water is on before proceeding with instrument startup.
Turn on POWER1 and POWER 2, then start the temperature controller (PTC-348MI) and computer.
Prepare sample in a flask, check concentration with UV/vis spectrometer, dilute if needed.
Fill the cuvette with pure solvent before starting the measurement process.
Choose 'Setup', configure parameters, insert solvent cuvette, and record a Baseline.
Replace solvent cuvette with sample, measure sample, and record multiple samples.
Close Specta Manager software and shut down the computer.
Turn off the PTC-348WI, Power 2 switch, and Power 1 switch.
Purge the instrument with N2 for 5 minutes after shutdown before turning off nitrogen flow.