P/N: 0302-1935D February 2010
Model V-630/650/660/670
Spectrophotometer
Hardware/Function Manual

Questions and Answers

  • A
    Angela CooperJul 31, 2025
    How to fix unstable performance at wavelengths shorter than 320 nm in Jasco V-630?
    • K
      Kevin WilliamsJul 31, 2025
      If the performance at wavelengths shorter than 320 nm is not stable on your JASCO Measuring Instruments, the D lamp might be off. Turn the D lamp on to resolve this issue.
  • V
    vjohnsonAug 5, 2025
    How to fix unstable performance at wavelengths longer than 400 nm in Jasco V-630?
    • R
      Ralph FernandezAug 5, 2025
      If the performance at wavelengths longer than 400 nm is not stable on your JASCO Measuring Instruments, the WI lamp might be off. Turn the WI lamp on.
  • A
    Ashley RayAug 9, 2025
    What to do if Jasco V-630 does not operate at all?
    • B
      Brian WrightAug 9, 2025
      If your JASCO Measuring Instruments device isn't operating, make sure the cable is securely reconnected. If the issue persists, check the fuse and replace it if it's blown. If a new fuse blows immediately, it indicates an electrical system fault, and you should contact your local JASCO distributor.
  • E
    etaylorAug 13, 2025
    What to do if the step height of the baseline is greater than 1% when the light source is changed from WI to D in Jasco Measuring Instruments?
    • M
      morgankimAug 13, 2025
      If, when the light source is changed from WI to D, the step height of the baseline is greater than 1% on your JASCO Measuring Instruments, try the following: 1. Set it to “ON” if set to “OFF”. 2. Re-adjust the mirror. 3. Set the standard cell holder in the correct position.
  • N
    Nicholas RamirezAug 17, 2025
    What to do if the baseline is curved greatly in Jasco Measuring Instruments?
    • A
      Antonio SchmidtAug 18, 2025
      If the baseline is curved greatly, in excess of 1% on your JASCO Measuring Instruments, ensure that it is set to “ON” if it is currently set to “OFF”.
  • C
    chrismorrisonAug 22, 2025
    What to do if WI lamp does not come on in Jasco V-630 Measuring Instruments?
    • T
      taylormirandaAug 22, 2025
      If the WI lamp does not come on in your JASCO Measuring Instruments, replace the lamp with a new one if it is burnt out.
  • H
    Helen MeyerAug 27, 2025
    What to do if D lamp does not come on in Jasco V-630?
    • S
      sharperAug 27, 2025
      If the D lamp does not come on in your JASCO Measuring Instruments, replace the lamp with a new one if it is burnt out.

Summary

Safety Considerations

Warning for Fuse

Details precautions for fuse replacement and grounding.

Warning for Ground

Explains the importance of proper grounding to avoid electric shock.

Connecting the Power Cable

Installation Requirements

Servicing

Limited Warranty

Unpacking and Installation

Installation and Connecting Cables

Connecting the Cable of iRM Type

Step-by-step guide for connecting cables for iRM type instruments.

Connecting the Cable of PC Type

Step-by-step guide for connecting cables for PC type instruments.

Specifications

Maintenance

Replacement of Consumables

Information on replacing consumable parts like fuses and lamps.

Replacement of Fuse

Detailed instructions and fuse specifications for replacement.

Lamp Replacement and Adjustment of the Light Source Mirror

Guides on replacing lamps and adjusting the light source mirror.

Lamp replacement

Step-by-step procedure for replacing the deuterium and halogen lamps.

Adjustment of iRM type light source mirror

Instructions for adjusting the light source mirror after lamp replacement.

Adjustment of PC type light source mirror

Instructions for adjusting the light source mirror for PC type instruments.

Troubleshooting

Jasco V-630 Specifications

General IconGeneral
TypeUV-Visible Spectrophotometer
Spectral Bandwidth1.5 nm
Wavelength Accuracy±0.3 nm
Wavelength Repeatability±0.1 nm
DetectorSilicon Photodiode
DisplayLCD
Wavelength Range190 to 1100 nm
Photometric Accuracy±0.002 Abs (at 1 Abs)
Photometric Repeatability±0.001 Abs
Stray Light<0.05% at 220 nm and 340 nm
Baseline Flatness±0.001 Abs
Light SourceDeuterium Lamp and Tungsten-Halogen lamps
MonochromatorCzerny-Turner mount
InterfaceUSB
Power RequirementsAC 100V to 240V, 50/60 Hz