P/N: 0302-1935D February 2010
Model V-630/650/660/670
Spectrophotometer
Hardware/Function Manual

Questions and Answers

  • D
    darlenebeckerAug 4, 2025
    How to fix unstable performance at wavelengths shorter than 320 nm in Jasco V-650?
    • N
      Nicholas MartinAug 5, 2025
      If the performance at wavelengths shorter than 320 nm is not stable on your JASCO Measuring Instruments, the D lamp might be off. Turn the D lamp on to resolve this issue.
  • W
    William MeyerAug 8, 2025
    How to fix unstable performance at wavelengths longer than 400 nm in Jasco Measuring Instruments?
    • N
      Nicholas JonesAug 9, 2025
      If the performance at wavelengths longer than 400 nm is not stable on your JASCO Measuring Instruments, the WI lamp might be off. Turn the WI lamp on.
  • M
    Melinda JacobsAug 13, 2025
    What to do if Jasco V-650 does not operate at all?
    • H
      Henry HarringtonAug 13, 2025
      If your JASCO Measuring Instruments device isn't operating, make sure the cable is securely reconnected. If the issue persists, check the fuse and replace it if it's blown. If a new fuse blows immediately, it indicates an electrical system fault, and you should contact your local JASCO distributor.
  • M
    Michele TaylorAug 17, 2025
    What to do if the step height of the baseline is greater than 1% when the light source is changed from WI to D in Jasco Measuring Instruments?
    • D
      Danny AyersAug 18, 2025
      If, when the light source is changed from WI to D, the step height of the baseline is greater than 1% on your JASCO Measuring Instruments, try the following: 1. Set it to “ON” if set to “OFF”. 2. Re-adjust the mirror. 3. Set the standard cell holder in the correct position.
  • B
    Brandon BowersAug 21, 2025
    What to do if the baseline is curved greatly in Jasco V-650 Measuring Instruments?
    • S
      Shelia GarciaAug 21, 2025
      If the baseline is curved greatly, in excess of 1% on your JASCO Measuring Instruments, ensure that it is set to “ON” if it is currently set to “OFF”.
  • D
    Deanna KnightAug 27, 2025
    What to do if WI lamp does not come on in Jasco V-650 Measuring Instruments?
    • T
      tammy46Aug 27, 2025
      If the WI lamp does not come on in your JASCO Measuring Instruments, replace the lamp with a new one if it is burnt out.
  • M
    Margaret Carter MDAug 30, 2025
    What to do if D lamp does not come on in Jasco Measuring Instruments?
    • M
      mullensylviaAug 31, 2025
      If the D lamp does not come on in your JASCO Measuring Instruments, replace the lamp with a new one if it is burnt out.

Summary

Instrument Safety Guidelines

Safety Symbols and Signal Words

Explains symbols like WARNING!, CAUTION, and Note used in the manual.

On-Instrument Warning Labels

Details warning labels on the instrument and their significance.

Fuse and Grounding Safety

Safety precautions related to fuses and proper electrical grounding.

Power Cable Connection

Instrument Installation Requirements

Unpacking and Initial Setup

Cable Connection Guide

Spectrophotometer Specifications

Instrument System Overview

Component Identification and Functions

Instrument Maintenance Procedures

Fuse Replacement Procedure

Instructions and fuse ratings for replacing fuses in V-630 and V-650/660/670 models.

Lamp Replacement Guidelines

Warnings and procedure for replacing halogen and deuterium lamps.

iRM Type Light Source Mirror Adjustment

Procedure for adjusting the light source mirror after lamp replacement for iRM type.

Halogen Lamp Mirror Adjustment

Steps to adjust the mirror for maximum signal with the halogen lamp.

Deuterium Lamp Mirror Adjustment

Steps to adjust the mirror for maximum signal with the deuterium lamp.

PC Type Light Source Mirror Adjustment

Procedure for adjusting the light source mirror after lamp replacement for PC type.

Troubleshooting Common Issues

Jasco V-650 Specifications

General IconGeneral
ModelV-650
CategoryMeasuring Instruments
TypeUV-Visible Spectrophotometer
Wavelength Range190 to 900 nm
Spectral Bandwidth1.5 nm
Wavelength Accuracy±0.3 nm
Wavelength Repeatability±0.1 nm
Resolution0.1 nm
DetectorSilicon Photodiode
DisplayLCD
Light SourceHalogen and Deuterium lamps
MonochromatorCzerny-Turner mount
InterfaceUSB
Power SupplyAC 100V to 240V, 50/60 Hz