P/N: 0302-1935D February 2010
Model V-630/650/660/670
Spectrophotometer
Hardware/Function Manual

Questions and Answers

  • D
    Debbie DanielsAug 4, 2025
    How to fix unstable performance at wavelengths shorter than 320 nm in Jasco Measuring Instruments?
    • L
      lmcdonaldAug 5, 2025
      If the performance at wavelengths shorter than 320 nm is not stable on your JASCO Measuring Instruments, the D lamp might be off. Turn the D lamp on to resolve this issue.
  • A
    Anne AllenAug 9, 2025
    How to fix unstable performance at wavelengths longer than 400 nm in Jasco Measuring Instruments?
    • J
      John KochAug 9, 2025
      If the performance at wavelengths longer than 400 nm is not stable on your JASCO Measuring Instruments, the WI lamp might be off. Turn the WI lamp on.
  • M
    Michelle HarrisAug 13, 2025
    What to do if Jasco Measuring Instruments does not operate at all?
    • R
      Rodney CarlsonAug 13, 2025
      If your JASCO Measuring Instruments device isn't operating, make sure the cable is securely reconnected. If the issue persists, check the fuse and replace it if it's blown. If a new fuse blows immediately, it indicates an electrical system fault, and you should contact your local JASCO distributor.
  • D
    Daniel MendozaAug 17, 2025
    What to do if the step height of the baseline is greater than 1% when the light source is changed from WI to D in Jasco V-670?
    • J
      jonathan20Aug 18, 2025
      If, when the light source is changed from WI to D, the step height of the baseline is greater than 1% on your JASCO Measuring Instruments, try the following: 1. Set it to “ON” if set to “OFF”. 2. Re-adjust the mirror. 3. Set the standard cell holder in the correct position.
  • D
    Debra JohnsonAug 22, 2025
    What to do if the baseline is curved greatly in Jasco V-670 Measuring Instruments?
    • K
      Kim MaxwellAug 22, 2025
      If the baseline is curved greatly, in excess of 1% on your JASCO Measuring Instruments, ensure that it is set to “ON” if it is currently set to “OFF”.
  • N
    nmackAug 27, 2025
    What to do if WI lamp does not come on in Jasco Measuring Instruments?
    • G
      Gina GomezAug 27, 2025
      If the WI lamp does not come on in your JASCO Measuring Instruments, replace the lamp with a new one if it is burnt out.
  • K
    Kristy LeeAug 30, 2025
    What to do if D lamp does not come on in Jasco V-670?
    • J
      John FrankAug 30, 2025
      If the D lamp does not come on in your JASCO Measuring Instruments, replace the lamp with a new one if it is burnt out.

Summary

Safety Considerations

Safety Symbols and Warnings

Explains safety symbols like WARNING, CAUTION, and Note, crucial for safe operation.

Warning Indication

Details how warnings are presented on the instrument for user safety.

Unpacking and Installation

Unpacking Procedures

Step-by-step guide for unpacking the main unit and standard accessories.

Installation and Cable Connections

Guidance on setting up the instrument and connecting cables for iRM/PC types.

Specifications

Maintenance

Consumable Parts Replacement

Guidance on replacing consumable parts like fuses and lamps, with part numbers.

Troubleshooting

Jasco V-670 Specifications

General IconGeneral
BrandJasco
ModelV-670
CategoryMeasuring Instruments
LanguageEnglish