0.1% (1690 nm CH
2
Br
2
50 mm cell used)
(spectral bandwidth: L8 nm)
Baseline stability 0.0003 Abs/hour
(Value obtained more than two hours after turning on the
light source, when the room temperature is stabilized,
wavelength: 250 nm, response: slow, and spectral
bandwidth: 2nm)
Baseline flatness
±0.0005 Abs
(Value obtained after baseline correction with a temperature
variation of less than 5°C, wavelength: 200 to 850 nm,
response : medium, spectral bandwidth: 2 nm, smoothing
processing and wavelength scanning speed: 400 nm/min,
spectral bandwidth: 8 nm in wavelength 850 to 2500 nm)
RMS noise 0.00003 Abs (0 Abs, wavelength: 500 nm, measurement
time 60 sec, response: medium, spectral bandwidth: 2 nm)
Power requirements
100, 115, 200, 220, 230, 240 V ±10 %, 145 VA
Dimensions and weight
270(H) ´ 460(W) ´ 602(D) mm (excluding protrusions)
28 kg
Note: Values given for wavelength accuracy, wavelength repeatability, photometric
accuracy, photometric repeatability, baseline flatness, and RMS noise are
those obtained more than one hour after the light source was turned on.