Operating / Service Manual
12
Leak test scope to verify no leaks
(pages 14-15)
Clean and disinfect scope
(page 17)
Check Scope Functions
1. Connect scope to light source
(pages 7, 10)
2. Look through scope to observe bright, clear image.
3. Move articulation lever and observe corresponding
movement of angulation section.
4. Inspect insertion tube and distal tip to verify both
are smooth and unblemished.
5. Test eyepiece focus adjustment.
6. Place scope near representative (anatomical)
target and verify colors observed by naked eye
under ambient light appear consistent with those
observed through scope using intended beroptic
illumination.
7. Verify high-intensity light exiting distal tip windows
is cool so as not to burn patient.
Scope is now ready for use.
Before Use:
DEVICE OPERATION - Functional Test Before Using
Ergo-Flex must be cleaned and disinfected
before each use.
Warning
!