3.7 Using a PED+ or PED-ER+ in the handheld mode
Personal dose is not accumulated in handheld mode
In the handheld mode, the PED+ and PED-ER+ measures the dose rate which is compensated for
the fact that it is no longer being worn on the body. Therefore, the PED+ should be held away from
the body whenever possible.
Use a PED+ or PED-ER+ in the handheld mode as follows:
1. On the Main menu > Handheld mode .
Attention: If the menu item is not available, enable the feature in DoseVision™.
2. Hold the PED so that the GM tube window is facing forwards or in the direction of the area being measured.
3. View the readings screen that shows the following:
Handheld mode icon
Indicates that PED is in the handheld mode.
Peak dose rate
Peak dose rate experienced since the PED was last reset using DoseVision™.
Dose rate bar graph
Bar graph representing the current dose rate.
Dose rate reading
Current dose rate measurement. The measurement is compensated for the device being worn away
from the upper body. The value is filtered to provide a stable reading.
Activity graph
Activity graph to help detect quick changes in radiation fields. The activity graph shows a rolling 30
second chart of the activity. Each bar represents the number of counts detected within the last second.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Johnson Matthey Tracerco PED Series and is the answer not in the manual?

Johnson Matthey Tracerco PED Series Specifications

General IconGeneral
BrandJohnson Matthey
ModelTracerco PED Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals