The battery level is so low that the radiological
performance of the PED cannot be guaranteed. The
PED shuts down 10 seconds after a battery critical
warning. To preserve the internal clock, do not
switch on the PED before recharging.
Calibration factors are outside normal range.
Return the PED to Tracerco.
Real-time clock failure. Could be the result of
completely flat battery.
Use DoseVision™ to reset the real-
time clock. After resetting the clock,
ensure that the PED time corresponds
to the computer time running
DoseVision™.
If no counts are recorded over 2 minutes (standard
range) or 8 minutes (extended range), the PED
assumes that the GM tube has failed. Also, for the
extended range only, after 4 minutes the dose rate
reading flashes to indicate low background and
warn of a possible failure of the GM tube.
Return the PED to Tracerco if normal
levels of background radiation were
expected.
Return the PED to Tracerco.
Memory is 95% full. When the memory is
completely full, the PED continues to measure and
display the accumulated dose, but date and time
information is not recorded.
Transfer data to DoseVision™.
The PED is not working correctly.
Return the PED to Tracerco.