Page 24
Instruction manual - G95
Kivex Biotec Ltd
7.2 Testing the battery
The charge holding level of the battery can be tested by:
Fully discharging the battery by leaving it switched on for five hours •
(the unit will turn off)
Fully recharging the battery by connecting to mains power for around two hours •
The charge light on HEAT key should be ON constantly.
The charge of the battery should hold for five hours, (depending on the ambient temperature)•
Note If the holding time for the battery is below five hours, please contact your local K-SYSTEMS
distributor or K-SYSTEMS.
8. The use of tubes or dishes
A selection of different tubes and dishes can be transported. Either tubes or dishes can be transported at
one time, not a combination of both.
Please contact your local distributor or K-SYSTEMS for details about the full range of warming blocks.
The empty compartment.
Note the small metal-stop in the back.