ORTHOPANTOMOGAPH™ OP 2D
Digital Panoramic X-ray Unit
Service Manual
ENGLISH
216143 rev.1
0.805.5058

Table of Contents

Question and Answer IconNeed help?

Do you have a question about the KaVo ORTHOPANTOMOGAPH OP 2D and is the answer not in the manual?

KaVo ORTHOPANTOMOGAPH OP 2D Specifications

General IconGeneral
Imaging TechnologyDigital
Exposure ParametersAdjustable
X-ray GeneratorHigh-frequency
ProgramsPanoramic
TypePanoramic X-ray
Image ResolutionHigh Resolution
Exposure TimeAdjustable

Summary

General Information

Service Precautions and Warnings

Details essential safety, electrical, and mechanical precautions for servicing the unit.

Yearly Maintenance

Outlines essential annual tests and inspections for ensuring correct and safe operation.

Circuit Boards

L1800, Generator Board

Covers the L1800 generator board, crucial for X-ray generation, including high voltage warnings.

Troubleshooting

Initial Checks

Outlines fundamental troubleshooting steps such as restarting, checking error codes, boards, and cables.

Common Issues

Addresses frequently encountered problems like unit not powering on, display issues, and control panel failures.

Error Codes

Lists and explains various system and user error codes that appear on the unit's main control panel.

Bad Quality Images

Discusses causes for poor image quality, including positioning, alignment, and calibration issues.

S2terminal and Service Commands

Using the S2terminal

Guide on how to open and use the S2terminal utility for unit testing and configuration.

Service Commands

Details various service commands for calibration, diagnostics, and unit configuration.

Replacing Parts

Replacing the Tubehead

Detailed procedure for replacing the tubehead, including required tools and steps.

Calibrating Tubehead and Checking Alignment

Post-replacement steps for calibrating the tubehead and verifying unit alignment.

Replacing the CCD Sensor

Steps for replacing the CCD sensor, including necessary post-replacement alignment.

Replacing Circuit Boards

General procedure for replacing circuit boards, including safety precautions and connection details.

Replacing the Z-Motor

Detailed procedure for replacing the z-motor, including safety precautions and access instructions.

Aligning the Unit

Checking and Aligning the CCD Sensor

Procedure for checking and aligning the CCD sensor for optimal image capture.

Adjusting Vertical Beam Position

Steps to adjust the vertical position of the X-ray beam for correct image framing.

Checking and Adjusting Collimator Position

Procedure to check and adjust the collimator for proper beam shaping and positioning.