EasyManuals Logo

Keithley 2400 Service Manual

Keithley 2400
137 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #20 background imageLoading...
Page #20 background image
2400-902-01 (G - Feb 2006)
Model 2400 Service Manual 1-8
Performance Verification
Taking the SourceMeter out of compliance
Verification measurements should not be made when the SourceMeter is in compliance. For
purposes of the verification tests, the SourceMeter can be taken out of compliance by going
into the edit mode and increasing the compliance limit.
NOTE Do not take the unit out of compliance by decreasing the source value or changing
the range. Always use the recommended range and source settings when performing
the verification tests.
Output voltage accuracy
Follow the steps below to verify that Model 2400 output voltage accuracy is within specified
limits. This test involves setting the output voltage to each full-range value and measuring the
voltages with a precision digital multimeter.
1. With the power off, connect the digital multimeter to the Model 2400 INPUT/OUTPUT
jacks, as shown in
Figure 1-1.
2. Select the multimeter DC volts measuring function.
NOTE The default voltage source protection value is 40V. Before testing the 200V range, set
the voltage source protection value to >200V. To do so, press CONFIG then
SOURCE V to access the CONFIGURE V-SOURCE menu, then select PROTEC
-
TION and set the limit value to >200V.
3. Press the Model 2400 SOURCE V key to source voltage, and make sure the source out-
put is turned on.
4. Verify output voltage accuracy for each of the voltages listed in Table 1-2. For each test
point:
Select the correct source range.
Set the Model 2400 output voltage to the indicated value.
Verify that the multimeter reading is within the limits given in the table.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2400 and is the answer not in the manual?

Keithley 2400 Specifications

General IconGeneral
Model2400
CategorySourceMeter
Voltage Source Range±200 V
Power Output20 W
Voltage Measurement Resolution100 nV
Current Measurement Resolution10 pA
Resolution5.5 digits
InterfacesGPIB, RS-232
Current Source Range±1A
Current Measurement Range±1A
Resistance Measurement Range2 Ω to 200 MΩ
DisplayVacuum Fluorescent Display