16-12 Return to Section Topics 2600S-901-01 Rev. C / January 2008
Section 16: Calibration Series 2600 System SourceMeter® Instruments Reference Manual
24) Typical values for the Model 2611/2612/2635/2636 2V range:
smua.source.calibratev(-2,-1e-10,-1e-4,-1.8,-1.805)
25) If this step is not on the CALA sense mode, send the measure calibration command
using the multimeter and Series 2600 readings, and range setting for the
parameters:
26) smua.measure.calibratev(-range,Z_rdg,DMM_Z_rdg, FS_rdg,DMM_FS_rdg)
Where: -range negative of the present calibration range
Z_rdg -zero 2600 measurement
DMM_Z_rdg -zero DMM measurement
FS_rdg -FS 2600 measurement
DMM_FS_rdg -FS DMM measurement
Typical Model 2601/2602 1V range values:
smua.measure.calibratev(-1,-1e-4,-1e-6,-0.89,-0.896)
27) Typical Model 2611/2612/2635/2636 2V range values:
smua.measure.calibratev(-2,-1e-4,-1e-6,-1.81,-1.805)
d. Be sure to complete steps a through v for all six voltage steps in Table 16-2, Table 16-3,
or Table 16-5 before continuing with current calibration.
e. Select auto polarity mode:
smua.cal.polarity = smua.CAL_AUTO

Table of Contents

Other manuals for Keithley 2601

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2601 and is the answer not in the manual?

Keithley 2601 Specifications

General IconGeneral
BrandKeithley
Model2601
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals