EasyManuals Logo

Keithley 2635 Reference Manual

Keithley 2635
562 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #318 background imageLoading...
Page #318 background image
12-58 Return to Section Topics 2600S-901-01 Rev. C / January 2008
Section 12: Instrument Control Library Series 2600 System SourceMeter® Instruments Reference Manual
Remarks This function generates and activates new calibration constants for the given range. The
positive and negative polarities of the instrument must be calibrated separately. Use a
positive value for range to calibrate the positive polarity and a negative value for range
to
calibrate the negative polarity.
Typically the two calibration points used will be near zero for calibration point 1 and 90%
of full scale for calibration point 2.
All four measurements (cp1measured, cp1reference, cp2measured, and
cp2reference) must be made with the active calibration set. Corruption of the
calibration constants may result if this is not heeded.
The new calibration constants will be activated immediately but they will not be
written to non-volatile storage. Use
smuX.cal.save to commit the new constants to
nonvolatile storage. The active calibration constants will stay in effect until the
instrument is power cycled or a calibration set is loaded from non-volatile storage
with the
smuX.cal.restore function.
This function will be disabled until a successful call to smuX.cal.unlock is made.
Details See Section 16 "Calibration".
Also see smuX.cal.restore, smuX.cal.save, smuX.makebuffer, smuX.source.calibrateY
smuX.measure.count
X = SMU channel (a or b)
Attribute Number of measurements performed when a measurement is requested.
Usage count = smuX.measure.count -- Reads measure count.
smuX.measure.count = count -- Writes measure count.
count Number of measurements.
Remarks This attribute controls the number of measurements taken any time a measurement is
requested. When using a reading buffer with a measure command, the count also
controls the number of readings to be stored.
The reset function sets the measure count to 1.
Details See “Triggering” in Section 10.
Also see smuX.measure.overlappedY, smuX.measure.Y
Example Sets measure count for SMU A:
smua.measure.count = 10
smuX.measure.delay
X= SMU channel (a or b)
(Models 2635 and 2636 only)
Attribute Controls the measurement delay.
Usage
mdelay = smuX.measure.delay -- Read the measure delay.
smuX.measure.delay = mdelay -- Write the measure delay.
Set mdelay to one of the following values:
0 or smuX.DELAY_OFF No delay.
-1 or smux.DELAY_AUTO Automatic delay value.
user_value Set user delay value.
Remarks This attribute allows for additional settling time before taking a measurement.
The default is smuX.DELAY_AUTO.
The smuX.DELAY_AUTO setting causes a current range-dependent delay to be inserted
when a current measurement function is called.
If smuX.measure.count is greater than 1, the measurement delay is only inserted
before the first measurement.
mdelay can be set to a specific user-defined value that sets the delay that is used
regardless of range.
Use smuX.source.delay when measuring your source (source read back) and
smuX.measure.delay when stepping voltage and measuring current.
Also see smuX.measure.delayfactor, smuX.source.delay
Example smua.measure.delay = 0.010 Sets the measurement delay to 10mS.

Table of Contents

Other manuals for Keithley 2635

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2635 and is the answer not in the manual?

Keithley 2635 Specifications

General IconGeneral
BrandKeithley
Model2635
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals